HATN-3000 is an automatic measuring instrument for ultraviolet / visible spectroscopy of water pollution test method. In the pretreatment step, alkaline potassium persulfate is injected into the sample, heated at 90°C for 15 minutes, and irradiated with UV light to decompose organic and inorganic nitrogen compounds into nitrate ions. After cooling, hydrochloric acid is added to lower the acidity, and ultraviolet absorbance is measured at 220nm to measure total nitrogen. At this time, it is possible to measure more accurately by measuring the absorbance of the disturbance factor at 275nm to calibrate the measurement value. HATN-3000 is a new product of KORBI which not only shortens measuring time by adopting patented technology and newly-designed oxidation reaction chamber, but also has sophisticated design and user convenience.
Characteristics
HATN-3000
Adoption of oxidation reaction chamber with improved efficiency
Application of multi-channel valve for quantitative distribution of samples and reagents
Built-in user-oriented operating program with enhanced security function
Built-in functions to enhance user convenience
Specification
-
- Measurement item
Total Nitrogen (TN)
-
- Measurement method
UV Absorption photometry
-
- Measurement range
0~100mg/L (Adjustable)
-
- Wavelength
220nm (275nm for compensation)
-
- Measurement interval
30 minutes (Adjustable)
-
- Detection limit
1mg/L F.S.
-
- Accuracy
±2% F.S.
-
- Reproducibility
±2% F.S.
-
- Linearity
±3% F.S.
-
- Zero/Span drift
±2% F.S.
-
- Oxidation method
UV persulfate oxidation
-
- Calibration
Auto/manual calibration by standard solution
-
- Sample/Reagent measuring method
Step motor pulse control
-
- Sample/Reagent distribution method
8-port distribution control method (Fixed quantity injection with multi-channel valve system)
-
- Operation system
Windows 7 Pro
-
- Output
DC 4~20mA, RS-232C
-
- Status display
Normal operation, on calibration, error
-
- Display
Color LED 10.4″ (Touch screen)
-
- Power
AC 220V, 50/60Hz
-
- Power consumption
Average 70W
-
- Voltage regulation
3% or less
-
- Data storage capacity
HDD 100G or more
-
- Data output
USB terminal
-
- Enclosure/Size
Steel, 1.5T (Except paint) / 645(W) x 439(D) x 1600(H) mm
-
- Operating humidity
RH 95% or less